AC WITHSTAND VOLTAGE
|
Output Rating
|
5KVAC / 50mA
|
5KVAC / 100mA
|
Range
|
Range
|
Resolution
|
Accuracy
|
Output Voltage, ACV
|
0 - 5000
|
1
|
± (2% of setting + 5V)
|
Output Frequency
|
50Hz / 60Hz ±0.1%, User Selection
|
Output Waveform
|
Sine Wave, Crest Factor = 1.3 - 1.5
|
Output Regulation
|
± (1% of output + 5V), From no load to full load and Low Line to High Line (combined regulation)
|
SETTINGS
|
HI and LO-Limit (Total) current, mA
|
0.000 - 9.999
|
0.001
|
± (2% of setting + 2 counts)
|
10.00 - 50.00 (for ESA-140)
|
0.01
|
10.00 - 100.00 (for ESA-150)
|
HI and LO-Limit (Real) current, mA
|
0.000 - 9.999
|
0.01
|
± (3% of setting + 50µA)
|
10.00 - 50.00 (for ESA-140)
|
0.1
|
10.00 - 99.99 (for ESA-150)
|
Ramp - Up Timer, second
|
0.1 - 999.9
|
0.1
|
± (0.1% of setting + 0.05 sec)
|
Ramp - Down Timer, second
|
0.0 - 999.9
|
Dwell Timer, second
|
0, 0.4 - 999.9
(0 = continuous)
|
Ground Continuity
|
Current: DC 0.1 A ± 0.01A, fixed
Max. Ground Resistance: 1.0 Ω ± 0.1Ω
|
Current Offset
|
0.000 - 50.00mA (PF=0.00 - 1.00)* (Total current + current offset ≤50mA) for ESA-140
0.000 - 99.99mA (PF=0.00 - 1.00)* (Total current + current offset ≤100mA) for ESA-150
|
Arc Detection
|
1 - 9 ranges (9 is the most sensitivity)
|
DC WITHSTAND VOLTAGE
|
Output Voltage, DCV
|
0 - 5000
|
1
|
± (2% of setting + 5V)
|
DC Output Ripple
|
< 4% (5KV/20mA at Resistive Load)
|
SETTINGS
|
HI and LO - Limit current, μA
|
0.0 - 999.9
|
0.1
|
± (2% of setting + 2 counts)
|
1000 - 20000
|
1
|
Ramp - Up Timer, second
|
0.4 - 999.9
|
0.1
|
± (0.1% of setting + 0.05 sec)
|
Ramp - Down Timer, second
|
0.0, 1.0 - 999.9
|
Dwell Timer, second
|
0, 0.3 - 999.9
(0 = continuous)
|
Ramp - HI current
|
> 20 mA peak maximum, ON / OFF User Selection
|
Charge - LO current
|
0.0 - 350.0 µA DC or Auto Set
|
Discharge Time
|
< 10mSec
|
Maximum Capacitive Load
DC Mode
|
1µF < 1KV 0.08µF < 4KV
|
0.75µF < 2KV 0.04µF < 5KV
|
0.5µF < 3KV
|
Ground Continuity
|
Current: DC 0.1 A ± 0.01A, fixed
Max. Ground Resistance: 1.0 Ω ± 0.1Ω
|
Arc Detection
|
1 - 9 ranges (9 is the most sensitivity)
|
INSULATION RESISTANCE
|
Output Voltage, DCV
|
30 - 1000
|
1
|
± (2% of setting + 2 counts)
|
Charging Current
|
Maximum > 20mApeak
|
SETTINGS
|
HI and LO - Limit resistance, MΩ
|
0.05 - 99.99 (HI - Limit: 0 = OFF)
|
0.01
|
0.05 - 999.9, ± (2% of setting + 2 counts)
|
100.0 - 999.9
|
0.1
|
1000 - 9999, ± (5% of setting + 2 counts)
|
1000 - 50000
|
1
|
10000 - 50000, ± (15% of setting + 2 counts)
|
Ramp - Up Timer, second
|
0.1 - 999.9
|
0.1
|
± (0.1% of setting + 0.05 sec)
|
Ramp - Down Timer, second
|
0.0 , 1.0 - 999.9
|
Dwell Timer, second
|
0, 0.5 - 999.9
(0 = continuous)
|
Delay Timer, second
|
0.5 - 999.9
|
Charge - LO current, μA
|
0.000 - 3.500 or Auto Set
|
GROUND BOND
|
Output AC Current, A
|
1.00 - 40.00
|
0.01
|
± (2 % of setting + 2 counts)
|
Output Voltage, Vac
|
3.00 - 8.00
|
0.01
|
± (2 % of setting + 3 counts)
|
Output Frequency, Hz
|
50Hz / 60Hz ± 0.1%, User Selection
|
Output Regulation
|
± (1% of output + 0.02A), Within maximum load limits, and over input voltage range
|
Maximum Loading
|
1.00 - 10.00A / 0 - 600mΩ, 10.01 - 30.00A / 0 - 200mΩ, 30.01 - 40.00A / 0 - 150mΩ
|
SETTINGS
|
Lead Resistance Offset, mΩ
|
0 - 200
|
1
|
± (2% of setting + 2 counts )
|
HI and LO - Limit Resistance, mΩ
|
0 - 150 (30.01 - 40.00A)
|
1
|
6.00 - 40A, ± (2% of setting + 2 counts )
|
0 - 200 (10.01 - 30.00A)
|
0 - 600 (6.00 - 10.00A)
|
0 - 600 (1.00 - 5.99A)
|
1.00 - 5.99A, ± (3% of setting + 3 counts )
|
Dwell Timer, second
|
0, 0.5 - 999.9
(0 = continuous)
|
0.1
|
± (0.1% of setting + 0.05 sec)
|
CONTINUITY TEST
|
Output Current
|
0.1A for 0 - 10.00Ω, 0.01A for 10.1 - 100.0Ω, 0.001A for 101 - 1000Ω, 0.0001A for 1001 - 10000, 0.1A is Max.
|
SETTINGS
|
Max and Min - Limit Resistance ,Ω
|
0.00 - 10.00
|
0.01
|
± (1% of setting + 3 counts)
|
10.1 - 100.0
|
0.1
|
101 - 1000
|
1
|
1001 - 10000
|
1
|
± (1% of setting + 10 counts)
|
Dwell Timer, second
|
0.0, 0.3 - 999.9
(0 = continuous)
|
0.1
|
± (0.1% of setting + 0.05 sec)
|
Resistance Offset, Ω
|
0.00 - 10.00
|
0.01
|
± (1 % of reading + 3 counts)
|
MEASUREMENT
|
|
Range
|
Resolution
|
Accuracy
|
Voltage, KV (AC / DC)
|
0.00 - 5.00
|
0.01
|
± (2 % of reading +1 count)
|
Voltage, Vdc (IR only)
|
0 - 1000
|
1
|
± (2 % of reading + 2 counts)
|
AC Current (Total), mA
|
0.000 - 3.500
|
0.001
|
± (2% of reading + 2 counts)
|
3.00 - 100.00
|
0.01
|
AC Current (Real), mA
|
0.000 - 9.999
|
0.001
|
± (3% of reading + 50µA)
All Ranges PF > 0.1, V > 250VAC
|
10.00 - 99.99
|
0.01
|
DC Current, μA
|
0.0 - 350.0
|
0.1
|
± (2% of reading + 2 counts)
|
DC Current, mA
|
0.300 - 3.500
|
0.001
|
3.00 - 20.00
|
0.01
|
AC Current, A (GB)
|
0.00 - 40.00
|
0.01
|
± (3 % of reading + 3 counts)
|
Resistance, MΩ (IR)
|
30 - 499V
|
500 - 1000V
|
|
30 - 499V
0.05 - 999.9, ± (7% of reading+2 counts)
500 - 1000V
0.05 - 999.9, ±(2% of reading+2 counts)
1000 - 9999, ±(5% of reading+2 counts)
10000 - 50000, ±(15% of reading+2 counts)
|
0.050 - 1.999
|
0.050 - 9.999
|
0.001
|
2.00 - 19.99
|
10.00 - 99.99
|
0.01
|
20.0 - 199.9
|
100.0 - 999.9
|
0.1
|
200 - 50000
|
1000 - 50000
|
1
|
Resistance, mΩ (GB)
|
0 - 600
|
1
|
1.00 - 2.99 A, ± (3 % of reading + 3 counts)
3.00 - 40.00 A, ± (2 % of reading + 2 counts)
|
Resistance, Ω (Continuity)
|
0.00 - 10.00
|
0.01
|
± (1 % of reading + 3 counts)
|
10.1 - 100.0
|
0.1
|
101 - 1000
|
1
|
1001 - 10000
|
1
|
± (1 % of reading + 10 counts)
|
GENERAL
|
Input Voltage AC
|
115/230Vac±15% Auto Range, 50/60Hz ± 5%, 5A/250Vac Slow-Blow for ESA-140, 10A/250Vac Slow-Blow for ESA-150
|
PLC Remote Control
|
Input : Test, Reset, Interlock, Recall File 1 through 3, Recall File 1 through 7(Option)
Output: Pass, Fail, Test-in-Process
|
Memory
|
10000 steps, allow the user create different memories and steps
|
TFT LCD
|
800 x 480 resolution digital TFT LCD, 9 ranges contrast setting
|
DualCHEK
|
5kVac/25mA and 25Aac/150mΩ for ESA-140
5kVac/50mA and 30Aac/150mΩ for ESA-150
|
Safety
|
Built-in Smart GFI circuit,GFI trip current 5.0mA max., HV shut down speed: <1mS
(on 50/60Hz and test under 1000V)
|
Hot Hipot Tests
|
To detect the line input voltage to produce a simultaneous sine wave of line power at hipot output
|
MyMenu
|
This is a menu that would be customizable by the user, allowing them to create a menu of their favorite most used functions and menu selection.
|
Interface
|
Standard USB & RS232 PC Control Card, Optional Ethernet, GPIB (IEEE-488.2), Multi-function Interface card (USB-A / RS-485 / RS-232 / BAR Code PS / 2 type)
|
Multinational language
|
The operate screen interface to provide selection different language including English / Traditional Chinese / Simplified Chinese
|
Alarm Volume Setting
|
Range: 0 - 9 ;0=OFF, 1 is softest volume, 9 is loudest volume.
|
Calibration
|
Software and adjustments are made through the front panel. Automatic Calibration alert function to signal
operator when re-calibration is due.
|
Environment
|
0 - 40ºC, 20 - 80%RH
|
Dimensions/Net Weight
|
430mm(W) × 133mm(H) × 500 mm(D) / 30Kg
|
OPTION
|
Opt.736 MATRIX SCANNER MODULE
|
High Voltage Rating
|
5KVAC/5KVDC
|
High Current Rating
|
40 A
|
Number of HV Channel
|
8
|
Number of HA Channel
|
8
|
Point to Point Continuity
|
Used the Scanner to reach point to point continuity test and this function is standard feature when built in scanner
|
Opt.767 RUN TEST MODULE
|
DUT POWER
|
AC Voltage
|
0 - 277.0V, Single phase unblance
|
Current
|
16A maximum continuous
|
Power Rating
|
4500W maximum
|
Short Circuit Protection
|
23Arms or Inrush Current 68Apeak, Response time RMS < 3 sec ; PEAK < 10uS
|
SETTINGS
|
HI and LO Limit AC Voltage, V
|
30.0 - 277.0
|
0.1
|
± (1.5% of setting + 0.2 V)
|
HI and LO Limit AC Current, A
|
0.00 - 16.00
|
0.01
|
± (2% of setting + 2 counts)
|
HI and LO Limit AC Power, W
|
0 - 4500
|
1
|
± (5% of setting + 3 counts)
|
HI and LO Limit Power Factor
|
0.000 - 1.000
|
0.001
|
± (8% of setting + 2 counts)
|
HI and LO Limit Leakage Current
|
0.00 - 10.00
Hi - Limit 0 = OFF
|
0.01
|
± (2% of setting + 2 counts)
|
Delay Time, second
|
0.2 - 999.9
|
0.1
|
± (0.1% + 0.05 sec)
|
Dwell Time, second
|
0, 0.1 - 999.9 (0 = continuous)
|
MEASUREMENT
|
|
Range
|
Resolution
|
Accuracy
|
Voltage, Vac
|
0.0 - 277.0
|
0.1
|
± (1.5% of reading + 2counts) 30 - 277V
|
Current, Aac
|
0.00 - 16.00
|
0.01
|
± (2% of reading + 2counts)
|
Power, Watts
|
0 - 4500
|
1
|
± (5% of reading + 3 counts)
|
Power, Factor
|
0.000 - 1.000
|
0.001
|
± (8% of reading + 2 counts)
|
Leakage Current, mA
|
0.00 - 10.00
|
0.01
|
± (2% of reading + 2 counts)
|
MD
|
Leakage current measuring resistor = 2kΩ ± 1%
|
Opt.768 TOUCH CURRENT TEST MODULE
|
DUT
|
DUT Input Power Rating
|
0-277V,AC@ 16Aac max
|
Current
|
16A maximum continuous
|
Short Circuit Protection
|
23Arms or Inrush Current 68Apeak, Response time RMS < 3 sec ; PEAK < 10uS
|
SETTINGS
|
Leakage current High/Low Limit (RMS), µA
|
Range
|
0.0-999.9µA (0=OFF),
|
Resolution
|
0.1µA
|
1000~10000uA
|
1µA
|
Leakage current High/Low Limit (peak), µA
|
Range
|
0.0-999.9µA (0=OFF),
|
Resolution
|
0.1µA
|
1000~10000uA
|
1µA
|
Dwell Time, second
|
0, 0.7-999.9 (0=continous)
|
0.1
|
± (0.1% + 0.05 sec)
|
Delay Time, second
|
0.5-999.9
|
Compliance Standards
|
A. UL544 Non Patient, UL484, IEC60598, UL1363,UL923, UL471, UL867, UL697
|
B. UL544 Patient Care
|
C. UL2601-1, IEC60601-1, EN60601-1
|
D. UL1563
|
E. IEC60990 Fig4 U2, IEC 60950-1, IEC60335-1, IEC60598-1, UL484, IEC60065, IEC61010,IEC60065
|
F. IEC60990 Fig5 U3, IEC60598-1
|
G. Basic measuring element 1k ohm of frequency check
|
MD A - G components
|
Accuracy: Resistance ± 1% Capacitance ± 5%
|
MD Voltage Limit
|
Maximum 30V peak or 30VDC
|
Probe settings
|
G-L, Ph-Pl, Ph-L (Use HV relay and HV terminal connector)
|
Internal Leakage
|
1. Internal Leakage current = 65uA
2. 277V applied to Ph max leakage current = 70uA.
|
External MD
|
User can add one extra MD for his application.
|
CURRENT MEASUREMENT
|
The leakage current is fitting range by leakage current Hi-limit setting value
|
Frequency Range
|
DC, 15Hz≤ F ≤1MHz
|
Leakage Current Range (RMS)
|
Auto Range
|
Range 1 - Range 6
|
0.0uA - 10.00mA
|
Resolution
|
0.1uA / 1uA / 0.01mA
|
Fixed Range > 6% of Range
|
Range 1 - Range 6
|
0.0uA - 10.00mA
|
Resolution
|
0.1uA / 1uA / 0.01mA
|
Fixed Range < 6% of Range
|
Range 2 - Range 6
|
0.0uA - 600uA
|
Resolution
|
0.1uA / 1uA / 0.01mA
|
Accuracy For Auto Range
|
Range
|
Mode
|
Frequency
|
Basic Accuracy
|
Range 1~5*1
|
AC+DC
|
DC,
|
±2% of reading + 3 counts
|
15Hz < f <100kHz
|
±2% of reading + 3 counts
|
100kHz < f < 1MHz
|
±5% of reading (>10.0uA)
|
AC only*2
|
15Hz < f <30Hz
|
±3% of reading + 5 counts
|
30Hz < f <100kHz
|
±2% of reading + 3 counts
|
100kHz < f < 1MHz
|
±5% of reading (>10.0uA)
|
DC only*3
|
DC
|
±2% of reading + 3 counts(>10.0uA)
|
Range 6*1
|
AC+DC
|
DC,
|
±5% of reading (>10.0uA)
|
15Hz < f <100kHz
|
AC only*2
|
15Hz < f <30Hz
|
30Hz < f <100kHz
|
DC only*3
|
DC
|
Accuracy For Fixed Range
|
Range
|
Mode
|
Frequency
|
Basic Accuracy(>6% of Range)
|
Additional Error(<6% of Range)
|
Range 1~5*1
|
AC+DC
|
DC,
|
±2% of reading + 3 counts
|
add (2% of reading + 0.2% of range)
|
15Hz < f <100kHz
|
±2% of reading + 3 counts
|
add (2% of reading + 0.2% of range)
|
100kHz < f < 1MHz
|
±5% of reading (>10.0uA)
|
add (2% of reading + 0.5% of range)
|
AC+DC
|
15Hz < f <30Hz
|
±3% of reading + 5 counts
|
add (2% of reading + 0.2% of range)
|
30Hz < f <100kHz
|
±2% of reading + 3 counts
|
add (2% of reading + 0.2% of range)
|
100kHz < f < 1MHz
|
±5% of reading (>10.0uA)
|
add (2% of reading + 0.5% of range)
|
DC only*3
|
DC
|
±2% of reading + 3 counts(>10.0mV)
|
add (2% of reading + 0.2% of range)
|
Range 6*1
|
AC+DC
|
DC,
|
±5% of reading (>10.0mV)
|
add (2% of reading + 0.2% of range)
|
15Hz < f <100kHz
|
AC only*2
|
15Hz < f <30Hz
|
30Hz < f <100kHz
|
DC only*3
|
DC
|
Leakage Voltage Range (Peak)
|
Auto Range
|
Range 1 - Range 6
|
0.0mV - 15.00V
|
Resolution
|
0.1mV / 1mV / 0.01V
|
Fixed Range > 6% of Range
|
Range 1 - Range 6
|
0.0mV - 15.00V
|
Resolution
|
0.1mV / 1mV / 0.01V
|
Fixed Range < 6% of Range
|
Range 2 - Range 6
|
0.0mV - 900mV
|
Resolution
|
0.1mV / 1mV / 0.01V
|
Accuracy For Auto Range
|
Range
|
Mode
|
Frequency
|
Basic Accuracy
|
Range 1~5*1
|
AC+DC
|
DC
|
±2% of reading + 2mV
|
15Hz < f < 1MHz
|
±10% of reading +2mV
|
AC only*2
|
15Hz < f < 1MHz
|
±10% of reading +2mV
|
Range 6*1
|
AC+DC
|
DC
|
±2% of reading + 3 counts
|
15Hz < f <100kHz
|
±10% of reading +2 counts
|
AC only*2
|
15Hz < f <100kHz
|
±10% of reading +2 counts
|
Accuracy For Fixed Range
|
Range
|
Mode
|
Frequency
|
Basic Accuracy(>6% of Range)
|
Additional Error(<6% of Range)
|
Range 1~5*1
|
AC+DC
|
DC,
|
±2% of reading + 2mV
|
add (2% of reading + 0.2% of range)
|
15Hz < f <100kHz
|
±10% of reading + 2mV
|
add (2% of reading + 0.2% of range)
|
100kHz < f < 1MHz
|
±10% of reading +2mV
|
add (2% of reading + 0.5% of range)
|
AC only*2
|
15Hz < f <100kHz
|
±10% of reading + 2mV
|
add (2% of reading + 0.2% of range)
|
100kHz < f < 1MHz
|
±10% of reading 2mV
|
add (2% of reading + 0.5% of range)
|
Range 6*1
|
AC+DC
|
DC,
|
±2% of reading +3 counts
|
add (2% of reading + 0.2% of range)
|
15Hz < f <100kHz
|
±10% of reading +2 counts
|
AC only*2
|
15Hz < f <100kHz
|
±10% of reading +2 counts
|
To explain with notes for leakage
|
*1.If the final measured signal is >range 5 then the maximum composite signal that can be measure is 28 volts peak。
If the final measured signal is ≤ range 5 then the maximum composite signal that can be measure is 12 volts peak。
|
*2.AC cutoff frequency for High Pass Filter is 15Hz on AC only mode。
|
*3.AC cutoff frequency for Low Pass Filter is 15Hz on DC only mode。
|
Leakage Imax Range
|
The specification is as same as leakage current(RMS)
|
The specification is as same as leakage current(Peak)
|
LINE VOLTAGE MEASUREMENT
|
Range
|
0.0-277.0 VAC
|
Resolution
|
0.1 V
|
Accuracy
|
± (1.5% of reading +0.2V), 30.0-277.0V
|
GENERAL
|
Continous Power Output selection (like Continuous Run) for both TCT and Run testing.
|
To create Continnuous parameter selection for both TCT and RUN testing.When continuous =ON, the power is not shut down between steps .But in the TCT mode, when the steps setting different line condition, DUT output will momentary power off 25ms then power on.
|
Opt.769 AC SOURCE 500VA
|
OUTPUT
|
Power
|
630VA and 500W Maximum
|
Voltage
|
0 - 150.0V / 0 - 277.0V
|
Current
|
4.20A / 2.10A
|